Specifications
Basic Information
Measurement method
Electrochemical Biosensor Technology
Enzyme type
GDH-FAD
Intended Use
Home testing or professional use to monitor the blood glucose (β-D-glucose) levels in fresh capillary whole blood
from fingertip and venous whole blood. Venous blood tests can only be measured by professionals.
Measuring Principle
Using the latest biosensor technology. It measures the glucose levels of the blood specimen by using a disposable
dry reagent strip which can produce an electrical current. And the current will transfer to the meter for measurement. The amount of the current is proportional to the amount of glucose
present in the blood sample. Test results are “plasma equivalent”.
Display mode
Reflective TN LCD
Key quantities & functions
3 keys (Mode, forward, backward)
Battery
2*CR2032
Operating environment
Temperature: 10℃~40℃ (50℉~104℉); Control solution:
20℃~30℃
Humidity: 80% R.H.
Storage/ Transport environment
Meter: Temperature: -40℃~55℃ / Humidity:
90% R.H.
Strip: Temperature: 4℃~30℃ / Humidity:
90% R.H.
Control: Temperature: 2℃~30℃ / Humidity:
90% R.H.
Product Life
Meter: 5 years
Strip: Shelf life: 24 months; Shelf life after opening: 180 days (Below 30℃)
Control: Shelf life: 24 months; Shelf life after opening: 90 days (Below 30℃)
Measurement range of glucose
10~600
mg/dl ( 0.5~33.3
mmol/l)
Control solution concentration range
Medium level: 120~164 mg/dl (6.6~9.2 mmol/l)
High level: 362~492 mg/dl (20.1~27.4 mmol/l)
Accuracy range of measurement
If glucose concentration < 100 mg/dl, Bias % (95% CI) ±15
mg/dl
else Bias % (95% CI) ±15% Ref. YSI 2300
System accuracy (ISO 15197)
Criterion A
95% of measured value fall within either ±15 (<100mg/dl) mg/dl or 15% (100 mg/dl)
of reference
Criterion B
99% of measured value within CEG zones A and B
Hematocrit Range
0%~70%
Measuring time
5 seconds
Sample volume
1.0 µl
Control solution volume
4 ml
Function
Unit Display
mg/dl and mmol/l
Time display
12h
Memory mode (Sets)
1000
Average display
Last 7 / 14 / 30 / 60 / 90 Days averaging
Talking function
No
Ketone warning / Voice indicator (Beep)
No
Insufficient sample volume detection
Yes
Meal Comment
Yes, Pre/Post Meal Marker
Before / After meals Average of blood glucose
No
Data-Link
AM100 PD-B: Yes (Including wired transmission and wireless transmission)
AM100 PD: Yes (Including wired transmission)
Package
Test Strip
vial
25 pcs/vial
50 pcs/vial
box
1 vial/ box, including instruction
2 vials/box, including instruction
Meter
box
Including 1 meter, 1 user’s guide, 1 Logbook, 1 Lancing device, 50 Lancets
Control Solution
Pouch
Including 1 control solution medium level, 1 control solution high level, control solution instruction (Packed in
the blood glucose meter pouch)